1.Walk me through how you'd troubleshoot an infusion pump that a nurse says keeps alarming for occlusion.
Why they ask
Occlusion alarms are one of the most common floor calls, and the cause is often not the pump. They want to see whether you start with the user and the setup before you open the case.
How to answer
- Talk to the nurse first: which channel, what fluid, which tubing set, and whether it happens on other pumps
- Check the tubing, clamps and cassette loading, since use error and set problems are a frequent cause
- Pull the pump, tag it out of service, and swap in a loaner so patient care isn't held up
- On the bench, pull the event log, run the manufacturer's pressure and flow checks against a flow analyzer, and compare to spec
- Document what you found, even if it was user error, and close the loop with the unit